"103B-0000" The automatic correction is not yet finished
Síntoma
"103B-0000" The automatic correction is not yet finished.
Solución
Situation/Error Cause
- Automatic correction has not been performed.
- Automatic correction was cancelled before it could finish.
- Automatic correction was not performed after updating the firmware to a version that required automatic correction to be performed again.
- The versions of VPanel and the machine's firmware do not match.
- Hold down the operation button on the built-in panel.This will clear the error.
- Download the latest versions of VPanel and the machine's firmware, and then install these versions.If you have already upgraded the version, proceed to the next step."DGSHAPE Corporation website (http://www.dgshape.com/)
- Perform automatic correction."P. 56 "Correcting the Milling Machine"