Article | Roland

DWX-53DC Dry Dental Milling Machine

Cargo
[101E-****][The % milling bur might be broken.]
Síntoma
[101E-****][The % milling bur might be broken.]
Solución
[101E-****][The % milling bur might be broken.]

[%] indicates a milling bur stocker number from [1] to [16].

Situation/Error Cause 1
• The milling bur is broken.
• The milling bur holder is out of position.

Procedure
1. Hold down the touch sensor on the built-in panel.
This will clear the error.
2. Check the status of the milling bur.
• When the milling bur is broken
Replace the milling bur with a new one.
• When the installation position of the milling bur holder is not appropriate
Install the milling bur holder in the correct position.
P. 40Step 3: Setting the Milling Bur
 
Situation/Error Cause 2
The milling conditions may be excessively strict.

Procedure
1. Hold down the touch sensor on the built-in panel.
This will clear the error.
2. Review the CAM settings and the shape specified in the CAD data.

If the Error Occurs Again
The collet may have worn out and its ability to retain the milling bur may have deteriorated. Replace the collet with a new one.
P. 100Replacing Consumable Parts
Modificado
September 14, 2022
ARRIBA