Base de conocimientos Article
Fresadora Dental DWX-52DCi de Cinco Ejes con Cambiador Automático de Discos
Cargo
Error Code "102E-0000"
Síntoma
Error Code "102E-0000" The spindle has collided with the rotary axis unit
Solución
Cause
The spindle has collided with the rotary axis unit.
Solution
The spindle has collided with the rotary axis unit.
Solution
- Turn off the power.
- Check the milling bur, the work piece, and the CAM settings as well as the shape specified in the milling data.
- Remove any pieces of cutting waste and foreign objects that have affixed to the adapter stocker.
- Turn on the power.
Modificado