DWX-4 4-Axis Dental Milling Machine

Cargo
Error Message 1022-0000
Síntoma
The milling bur is not found.
Solución
Situation/error cause
The milling bur has not been set or it may have been mounted on an incorrect stocker number.

Action
  1. Set the milling bur in the correct position.
  2. If the error occurred during milling
    • Close the front cover, and press the built-in panel's operation button. Resume milling.
  3. If the error occurred outside of milling
    • Close the front cover, and then press and hold the built-in panel's operation button. This will clear the error.


Situation/error cause
The ATC magazine or the milling bur might be dirty.

Action
  1. Clean the ATC magazine and the milling bur.
  1. If the error occurred during milling
    • Close the front cover, and press the built-in panel's operation button. Resume milling.
  2. If the error occurred outside of milling
    • Close the front cover, and then press and hold the built-in panel's operation button. This will clear the error.


Situation/error cause
There is a possibility that the ATC magazine is out of position.

Action
If the error occurs again even after you perform the above operations, contact Roland DGA Corporation.



Situation/error cause
There is a possibility that the spindle unit is defective.

Action
If the error occurs again even after you perform the above operations, replace the spindle unit with a new one. After you replace the spindle unit, correct the milling machine.
P. 49 "Replacing the Spindle Unit" , P. 39 "Milling Machine Correction"
Modificado
May 05, 2020