"101D-000*" The % milling bur cannot be released
Síntoma
"101D-000*" The % milling bur cannot be released.
The milling bur stocker number (1 to 7) is displayed for %.
Solución
Situation/Error Cause 1
Situation/Error Cause 2
If the Error Occurs Again
- The ATC magazine is dirty
- The inside of the collet is dirty
- Follow the instructions in the VPanel window to open the collet and remove the milling bur.
- Clean the ATC magazine. "P. 51 "Cleaning after Milling Finishes"
- Perform collet maintenance."P. 44 "Collet Maintenance, Cleaning of Dummy Pin, and Checking of Coolant Flow Rate"
Situation/Error Cause 2
- The stocker is out of position
- Follow the instructions in the VPanel window to open the collet and remove the milling bur.
- Perform automatic correction. "P. 56 "Correcting the Milling Machine"
If the Error Occurs Again
- The collet is deformed.
- Replace the collet."P. 83 "Replacing the Collet"
- Perform automatic correction."P. 56 "Correcting the Milling Machine"