Article | Roland
Knowledge Base Article

DWX-52DCi 5-Axis Dental Milling Machine With Automatic Disc Changer

Title
Error Code "1022-****"
Summary
Error Code "1022-****" The % milling bur was not found.
Details
Note: % indicates a milling bur stocker number from 1 to 15.

Cause #1
The milling bur has not been set or it may have been set on an incorrect stocker number.

Solution
Set the milling bur in the correct position.

If the error occurred during milling:
Press the operation button on the built-in panel.
Milling will resume.

If the error occurred while the machine was on standby:
Hold down the operation button on the built-in panel.
This will clear the error.


Cause #2
There is a possibility that the ATC magazine is out of position. 

Solution 
Perform automatic correction.
Refer to User Manual P. 65 "Correcting the Milling Machine".


If the Error Occurs Again
The collet may have worn out. Replace the collet. If the error occurs again even after you replace the collet, replace the spindle unit.
Refer to User Manual P.62 "Situations Requiring Part Replacement".
Modified
April 10, 2020