Error Code "1017-0001"
Summary
Error Code "1017-0001" The front cover or the magazine cover was opened.
Details
Cause
The front cover or the magazine cover was opened while the disk changer was operating. For safety, the machine comes to an emergency stop if a cover is opened while the disk changer is operating. Do not open the covers while the disk changer is operating.

Solution

To continue milling:
Press the operation button on the built-in panel.
Milling will resume.

To abort milling:
Hold down the operation button on the built-in panel.
Modified
April 10, 2020