DWX-52D 5-Axis Dental Milling Machine

Title
Error Code "1023-0000 to 1028-0000"
Summary
Error Code "1023-0000 to 1028-0000" Milling data error.
Details
Cause #1
There may be a problem with the milling data received from the computer. 

Solution
  1. Hold down the operation button on the built-in panel.
    1. This will clear the error. 
  2. Review the CAM settings and the shape specified in the CAD data.
If there are no problems with the milling data:
Restart the computer and then perform milling again.


Cause #2
When using CAM to select the stocker number, the stocker number set as the second or third milling bur in Intelligent Tool Control has been selected.

Solution
When Using CAM to select the stocker number, do not select the stocker number set as the second or third milling bur in Intelligent Tool Control.
Refer to User Manual P. 55 "Automatically Switching Out the Worn Milling Bur (Intelligent Tool Control)"