"102B-0000" The spindle motor temperature is too high
Síntoma
"102B-0000" The spindle motor temperature is too high.
Solución
Situation/Error Cause 1
Situation/Error Cause 2
If the Error Occurs Again
There is a possibility that the spindle unit is defective. To replace the spindle unit, contact Roland DGA Corporation.
- The milling bur is worn
- A workpiece that cannot be milled by the machine is being used
- The milling conditions are too strict
- Turn off the power.
- Check the milling bur, the workpiece, and the CAM settings as well as the shape specified in the milling data.
- Allow the machine to rest for some time before turning the power on.The motor may have overheated.
Situation/Error Cause 2
- Milling waste has adhered to the spindle.
- Perform spindle run-in (long). "P. 95 "Spindle Run-In (Long)"
If the Error Occurs Again
There is a possibility that the spindle unit is defective. To replace the spindle unit, contact Roland DGA Corporation.