Article | Roland

DWX-4W Wet Dental Mill

Cargo
Error Message 1033-0000
Síntoma
The coolant has run out.
Solución
Situation/error cause
  • There may be no coolant in the tank.
  • There may not be enough coolant in the tank.
Action
  1. Supply the appropriate amount of coolant.
  1. If the error occurred during milling
    • Press the built-in panel's operation button. Resume milling.
  2. If the error occurred outside of milling
    • Press and hold the built-in panel's operation but-ton. This will clear the error.
  3. If the error occurs again even after you perform the above operations, replace all the coolant.

Roland DGA Store

Coolant Additive for DWX Wet Mills - SKU# ZAW-1000D


Chelating Agent for DWX Wet Mills -  SKU# ZCH-250D


 

Situation/error cause
The filter or the lines may be clogged.

Action
Clean the inside of the coolant tank. If the filter is dirty, replace it.
P. 49 "Replacing the Coolant Tank Filter"

Roland DGA Store

Coolant Filter - DWX-4W - SKU# ZF-CLD


 

Situation/error cause
The coolant pump or sensor may be defective.

Action
If the error occurs again even after you perform the above operations, turn off the power and contact Roland DGA Corporation.