"1023-0000 to 1028-0000" Milling data error
Síntoma
"1023-0000 to 1028-0000" Milling data error. • 1023-0000: Milling data error. The number of parameters is incorrect. • 1024-0000: Milling data error. The parameter is out of range. • 1025-0000: Milling data error. A wrong command is detected. • 1026-0000: Milling data error. The address is not defined. • 1027-0000: Milling data error. The parameter is not defined. • 1028-0000: Milling data error. The operation cannot be executed
Solución
Situation/Error Cause
  • There is a problem with the milling data received from the computer.
  • The data transfer has failed due to the computer being under a high load.
Procedure
  1. Hold down the operation button on the built-in panel. This will clear the error.
  2. Check the milling data.
If necessary check the CAM settings and the shape in the CAD data.

If there are no problems with the milling data
  1. Exit any applications that are not being used.
  2. Restart the computer.
  3. Perform milling again