"101F-000*" The % milling bur chucking has slipped out
Síntoma
"101F-000*" The % milling bur chucking has slipped out. The milling bur stocker number (1 to 7) is displayed for %
Solución
Situation/Error Cause 1 Procedure
  1. Hold down the operation button on the built-in panel. This will clear the error.
  2. Replace the milling bur with a new one.
  3. Perform collet maintenance. "P. 44 "Collet Maintenance, Cleaning of Dummy Pin, and Checking of Coolant Flow Rate


Situation/Error Cause 2
  • The milling conditions are too strict.
Procedure
  1. Hold down the operation button on the built-in panel. This will clear the error.
  2. Review the CAM settings and the shape specified in the CAD data.


If the Error Occurs Again
  • The collet has worn out, leading to a decrease in its retention capabilities.
Procedure
  1. Replace the collet. "P. 83 "Replacing the Collet"
  2. Perform automatic correction. "P. 56 "Correcting the Milling Machine"
Modificado