[103D-0000][Milling data error. The milling bur cannot reach the milling position.]

Situation/Error Cause
If the milling bur is too short, or if the angle of the A axis and B axis are too large, the milling bur may not reach the milling position in the Z direction.

Procedure
1. Hold down the touch sensor on the built-in panel.
This will clear the error.
2. Review the milling bur length and the position of the milling bur holder.
P. 40Step 3: Setting the Milling Bur
3. Review the CAM settings and decrease the angles of the A axis and the B axis.
Modified
September 14, 2022